3300/2400 Series Imaging Detector Systems
3300/2400 Technical Description June 1999

Shown: Model 3300 Series
Detector Head (UHV type), 2401B Position Analyzer
MAJOR FEATURES
-
Ultimate, single-event-counting sensitivity, detects individual charged-particles and energetic photons including electrons, positrons, ions, energetic neutrals, EUV and soft x-rays. Open-face detector head specifically designed for use in high-vacuum environments.
-
Position-sensitive, 1-D and X-Y, with up to 1024 x 1024 channels (pixel) spatial digitization. Better than 60 micron FWHM spatial resolution.
-
Displays XY event data on detector in real-time on conventional oscilloscope in X-Y mode.
-
Extremely low detector noise combined with single-event, pulse-counting detection means higher S/N ratio and better data in detector-noise limited applications.
-
Capture more signal with 18 mm, 25 mm, 40mm, 75 mm and 80 x 100 mm active collection areas to match field-of-view. Fully UHV-compatible.
-
Fast data acquisition with count rates from less than one count/second to 100,000 counts/sec (other models up to 1 million counts/sec).
-
Optional 100 psec FWHM single-event time-resolution combined with spatial imaging available using MCP and external timing electronics.
-
Easy-to-use, complete systems to meet application needs including detector heads, readout electronics, powerful data collection/display hardware and software, options and accessories.
-
Proven, dependable product design and support. High-quality materials, design and fabrication.
-
Used by satisfied customers worldwide in a wide range of applications including XPS, EELS, LEED, SIMS, Imaging TOF MS, soft x-ray, EUV.
INTRODUCTION
The efficient detection of particle and photon flux has represented a major instrumentation challenge since the beginning of science. Today, scientific, analytical and industrial research demands even higher detector performance as many efforts focus on observation of inherently weak signal-generating processes, smaller areas, spatially-resolved analysis and the need to gather data quickly and minimize exposure and of damage-sensitive sample materials. Simply, there is the need to make more spatially-resolved measurements, more accurately, faster and at lower levels than ever before.
Quantar Technology's 3300/2400 Series Multichannel/ Multipixel ImagingTM Detector Systems are position-sensitive, quantum-limited, charged-particle and photon detectors designed primarily for these kinds of scientific applications.
These high performance detectors are ideal solutions in a wide range of 1-D parallel multichannel spectroscopy and 2-D X-Y imaging applications.
Complete systems include MCP/RAE detector heads
(3300
Series Sensors)
, signal processing and readout electronics
(2400/2500
Series Position Analyzers)
and PC-based data collection and display hardware and software
(2200
Series
Software and associated hardware boards and other
multiparameter data systems).
SYSTEM PERFORMANCE
3300 Series MCP/RAE Sensors (detector heads) utilize MCP (microchannel plate) imaging electron multipliers followed by a charge-division position-encoding anode. These detectors provide true, single-event, position-sensitive detection of electrons, positrons, positive and negative ions, neutral particles and EUV and soft X-ray photons (10-200 nm) in vacuum environments. Higher energy X-rays (>2 KeV) can also be detected directly by MCP's with decreasing efficiency at higher energies.
Combined with extremely-low detector MCP background dark count (typically less than 10 counts/sec over entire active area for 25mm MCP's), this approach offers extremely high signal-to-detector-background ratio measurements, especially useful in relatively low count-rate applications, where signals of interest would otherwise be obscured by noise.
Detector heads are fabricated using precision alumina ceramic and gold-plated stainless steel for maximum UHV compatibility (vacuum pressures of better than 10-6 are typically required for operation of these detectors).
For optical photon detection, the 2600 Series Mepsicron-IItmImaging Detector Systems are also available for true single-photon detection and imaging in the 180-900 nm range (UV-VIS-near IR). These are permanently vacuum-sealed, intensifier-like imager tubes and incorporate an optically-sensitive photocathode on a transparent faceplate. The remainder of the system is identical to that described above.
These optical detectors are an
excellent choice for ultra-low count-rate optical applications such as
Raman scattering,
fluorescence and luminescence spectroscopy and imaging,
and astronomy. Using external timing electronics, both
single-photon time-resolved and imaging capabilities can be
simultaneously combined using this unique detector, a
powerful and unusual capability. For more information on
this optical version, see the 2600 Series Mepsicron-II tm
Technical Description.
SYSTEM OPERATION
In operation, single incoming events (particles or photons) impact the MCP surface and result in an electron cascade, the gain depending on the number and type of MCP's, the bias voltage and the MCP configuration. The MCP's operate in a gain-saturated mode to ensure a relatively constant gain from event-to-event to optimize position determination. Detection efficiency for various types of particles and photons is determined primarily by the MCP surface response or by applied enhancement coatings.
The resulting charge packet from the MCP stage is electrostatically focused on the charge-division anode. Four pulse signals are processed by separate charge-sensitive preamplifiers, shaper amplifiers and discriminators. The X and Y spatial coordinates of each detected event, corresponding to a single incoming particle or photon, are then computed from the ratio of charge pulse amplitudes. The X and Y position is output in analog format for real-time display on an analog CRT XY monitor (such as a 2-channel oscilloscope operated in X-Y mode) and in digitized format (see options) for use in digital data collection systems.
By accumulating single events for a
user-determined period
of time, spectra and images are developed from the
integration of these single events.
SPATIAL RESOLUTION
3300 Series Sensor heads are available in several configurations with circular active diameters of 18 mm, 25 mm, 40 mm, 75 mm and a nominal rectangular size of 80 x 100 mm (74 x 93 mm actual useable area). Custom sizes are available to meet specialized application requirements. Special features are available, such as 6 mm diameter clear center holes through the entire detector to accommodate coaxial electron or ion guns.
Standard and high spatial-resolution versions are available. Standard resolution versions achieve 100 resolution elements (1/100 of active dimension) across each axis. High-resolution versions achieve 400 resolution elements (1/400 of active dimension).
These versions differ primarily in the number of MCP stages and the corresponding electron gain. In these systems, FWHM spatial resolution (position blur or point-spread distribution for the reported position of a statistically large number of events) is the result of event-to-event position jitter caused by slight variations in the computed spatial position of successively imaged events. For fixed noise, higher MCP gain reduces the effect of this position jitter, improving fractional position resolution.
Measured spatial resolution can also be
affected by the
spatial sampling intervals of the digitized output. Digital
output options are available with either 8-bit (256 channels),
9-bit (512 channels) or 10-bit (1024 channels) equally
spaced along the X and Y axis. Generally, the number of
digital channels (and therefore the digital output option) should be
chosen to provide at least 2X finer spatial
sampling than the required spatial resolution. For example,
for a desired spatial resolution of 100 resolution elements
across the active detector dimension, at least 200 points,
preferably more, should be digitally sampled. Higher ADC
resolution options, however, reduce the maximum
achievable count rate due to longer system dead time.
COUNT RATE CAPABILITY
Each detected event is processed separately and occupies a fixed amount of signal processing time. Only one event can be processed simultaneously. This fixed time period is required for position calculation and preamp/shaper baseline recovery to ensure accurate spatial position calculations of subsequently-occurring event.
The basic paralyzable dead time of the Model 2401B Position Analyzer, including the preamplifiers, is 4 microseconds per detected event, and applies to both the analog-only output version and the 8-bit digitized output option. This dead time is increased to 6 microseconds for 9-bit digitized output, and to 10 microseconds for 10-bit digitized output (the added time is non-paralyzable). Digitization is performed by fast, highly-linear 130 MHz clock rate, Wilkinson-type ADC's.
A fast "look-ahead" channel is incorporated for pulse-pile-up detection and event rejection. If a second event is received by the charge-sensitive amplifier within the dead time (in µsec) of a prior event, either one or both of the events are vetoed from further processing to ensure optimum spatial position calculation accuracy. A separate RATE output signal (500 nsec dead time) is derived from this channel for use in dead time corrections of count rate and to monitor input count rate. A front panel meter monitors this RATE signal, MCP gain and dead-time loss percentage.
For perfectly time-periodic input
counts, the 4 µsec dead
time translates to a maximum output count rate of 250,000
cts/sec. For Poisson-distibuted time-of-arrival statistics
encountered in typical applications, this maximum output
rate is reduced due to the arrival of closely-time-spaced
events that are lost from processing and the effect on
maximum output count rate is shown below.

DYNAMIC COUNTING RANGE
In a single-event-counting detector,
the time-integrated
dynamic counting range is limited only by available digital
storage memory. Each event count is stored as a single
count, with no crosstalk or other interaction with adjacent
channels or pixels. This characteristic is particularly advantageous
when detecting spectra or images with
spectral or spatial features of significantly differing
intensities. In the 2200 Series data system, data memory
counting range is typically 16 bits deep (65,000 counts per
pixel) but can be selected to be 32 bits deep (4,000,000,000
counts).
TIME-RESOLVED CAPABILITY
There are several methods to obtain time-resolved event data with these detectors, depending on the application.
The Model 2401B Position Analyzer can be electronically gated to exclude unwanted events from the output data using an external TTL-level signal on the VETO GATE input, with a time-resolution of approximately 500 nsec.
Using conventional external
single-event pulse timing
equipment (amplifier/CFD/TAC), event position data can
be "time-tagged" using the MCP bias recharge pulse signal
relative to a reference pulse from, for example, a laser
source. Digitized representations of this time can then be
written to a data system together with the event XY position
data or used in an external coincidence system to exclude
unwanted events. Time resolution of better than 100 psec
FWHM has been achieved using this method.
DATA COLLECTION AND DISPLAY
Real-time event positions can be readily displayed on an analog XY CRT monitor (oscilloscope in XY mode) using the 0-5V analog X and Y axis position pulses. Each event is displayed as a momentary bright dot on the CRT in its proper image position. This display mode is extremely useful for real-time spectral and image alignment and diagnostics.
These analog position pulses can also be displayed using a one or two-dimensional multichannel analyzer (MCA) with an integral pulse-height analyzer ADC.
Units equipped with digitalized output options produce TTL-level, positive-true, 8 or 9 or 10 bit position addresses for both X and Y axis, accompanied by a digital strobe signal. These signals can be then read directly by a computerized data system. Alternatively, the digital position data can be captured by an auto-incrementing (add-one), dual-ported histogramming memory. This memory has a dedicated memory channel for each spatial position in the spectra or image. Under software control, the memory can start and maintain an accumulation and data can be simultaneously read out into a PC-based data system and the computer display screen updated. This histogramming memory approach avoids PC-bus limitations and enables significantly higher data rates to be realized than can be achieved with direct readout into a PC.
Quantar Technology offers several
fully-integrated data
system configurations with histogramming memories of
different sizes to match application requirements, and both
1-D spectral and 2-D imaging DOS-based software. See
2200 Series Data Sheet for additional detail. Also available
are PC-based, 3-parameter Multiparameter Data Systems for
X-Y-T data collection and display.
ACCESSORIES
Quantar can provide HV bias power
supplies and divider
networks for MCP biasing, time-resolving subsystems,
special cables, vacuum flanges, isolated mounting ring
options to enable mounting accelerating, collection or
retarding grids in front of the input MCP and other
accessories upon request.
APPLICATIONS
Quantar Technology's detector systems are being successfully used worldwide and have proven useful in a wide range of spectroscopic and imaging applications in scientific and analytical instrumentation fields including:
- Parallel multichannel electron spectroscopy, XPS, EELS
- Secondary-ion imaging mass spectrometers (SIMS),
- Imaging TOF-SIMS
- Sector-type mass spectrometers
- LEED, ESDIAD (electron-stimulated ion angle) and SARIS (angle-resolved ion TOF MS) imaging
- Soft x-ray fluorescence multichannel spectroscopy
- Raman multichannel spectroscopy and imaging
- Fluorescence and luminescence spectroscopy/imaging
- Space science and astronomy
- Neutron imaging (with scintillator)
- Biomedical applications
Quantar Technology invites inquires regarding possible applications.
| |